MDCG 2021-1 Rev.1
Guidance on harmonised administrative practices and alternative technical solutions until EUDAMED is fully functional
Disclaimer: This document is an interactive version of the original MDCG document. We will keep it up-to-date.
This document has been endorsed by the Medical Device Coordination Group (MDCG) established by Article 103 of Regulation (EU) 2017/745. The MDCG is composed of representatives of all Member States and it is chaired by a representative of the European Commission.
MDCG 2021-1 Rev.1 changes
MDCG 2021-1 Rev.1 changes
|
|
---|---|
Third paragraph
|
Sentence deleted: “Parties should also note that, in principle and unless otherwise concluded by the MDCG, the reporting obligations with regard to EUDAMED apply to all information generated and collected under the MDR from its date ofapplication (e.g. UDI, certificates) and therefore must be carried out as soon as EUDAMED becomes fully functional”
|
INTRODUCTION
Article 33 of Regulation (EU) 2017/745 on medical devices (1) (MDR) requires the Commission to set up a European database on medical devices (‘EUDAMED’). According to Article 33(2) of the MDR, Eudamed will be composed of six different electronic systems (so called ‘modules’), which facilitate the collation and processing of information under the MDR regarding the registration of relevant economic operators (actor registration), devices and systems and procedure packs (UDI), notified bodies & certificates, certain aspects of conformity assessment, clinical investigations, vigilance and market surveillance as well as post-market surveillance.
On 30 October 2019, the Commission published a notice (2) by which it concluded that the full functionality of EUDAMED requires the availability and full operation of all six modules, both individually and jointly. The notice foresees the launch of EUDAMED for May 2022, which correlates with the date of application of Regulation (EU) 2017/746 on in vitro diagnostic medical devices (3) (IVDR). In this regard, it is important to note that the official launch of EUDAMED in May 2022 does not affect the date of application of the MDR on 26 May 2021.
Article 123(3)(d) MDR addresses the possibility that EUDAMED is not fully functional on the date of application of the MDR (26 May 2021). Accordingly, the obligations and requirements in the MDR that relate to EUDAMED shall apply from the date corresponding to six months after the date of publication of the notice referred to in Article 34(3) – notice of full functionality of Eudamed. Until EUDAMED is fully functional, the MDR stipulates that the corresponding provisions of Directives 90/385/EEC (4) and 93/42/EEC (5) shall continue to apply for the purpose of meeting the obligations laid down in the provisions of Article 123(3)(d) regarding the exchange of information.
In addition, Article 123(3)(e) MDR clarifies that Article 29(4) MDR on the registration of devices, and Article 56(5) MDR on the registration of certificates, start to apply 24 months after the date of publication of the notice referred to in Article 34(3) MDR.
Article 122 is outside scope of this guidance.
SCOPE
This document provides guidance to Member States and other relevant parties on the application of certain MDR provisions during the absence of EUDAMED. To that end, this guidance intends to describe harmonised administrative practices and alternative technical solutions for the exchange of information until EUDAMED becomes fully functional.
The proposed practices and solutions aim to enable Member States and other relevant parties to meet their obligations under the MDR effectively while minimising any potential additional burden on the parties concerned. This guidance addresses in particular cases where the exchange of information would be difficult, or even not possible, to achieve based on the corresponding provisions of Directives 90/385/EEC and 93/42/EEC. In doing so, this guidance takes into account the decision of the Medical Device Coordination Group in its meeting of 12 March 2020 where the group agreed that the Commission makes available to Member States each EUDAMED module as soon as it is operational. This approach also has an impact on the means by which relevant information collected under the MDR will be made available to the public, which shall take place on a gradual basis.
The proposed practices and solutions set out in this document do not affect the general obligations of the parties to comply with the requirements under the MDR, including those contained in the provisions referred to in Article 123(3) letters (d) and (e) MDR.
Whenever this guidance makes reference to CircaBC (6) as alternative solution, the Commission and other relevant parties should endeavour to make use of already existing CircaBC directories to the extent that this is possible and appropriate.
Parties should also take note of the MDCG Position Paper on the use of the EUDAMED actor registration module and of the Single Registration Number (SRN) in the Member States. (7)
Article | Provisions related to the use of EUDAMED (as referred to in Art. 123(3)(d) MDR) | Alternative solutions to submit and/or exchange information (as required under the MDR) | Responsible actor(s) |
---|---|---|---|
Article 29: |
| Paragraphs 1-4: Nevertheless, manufacturers should refer to the national provisions in Member States establishing product registration schemes. Manufacturers should note that the obligation of UDI assignment (Basic UDI and UDI-DI) to a device applies from 26 May 2021 (Art. 27(3) MDR). Labelling requirements apply gradually, starting from 26 May 2021 (Art. 123(3)(f) MDR). | Manufacturers |
Article 31: |
[…] | Paragraphs 1-3: Nevertheless, manufacturers, authorised representatives and importers should refer to the national provisions in Member States. | Economic operators (registration) Member States |
Article 32: |
[…] | Paragraphs 1: As soon as the functionality is available in Eudamed, the system may be used for the upload of the SSCP even before the notice of full functionality of Eudamed has been published. Paragraph 2: | Manufacturers Notified Bodies |
Article 33: | […] 2. Eudamed shall include the following electronic systems: (a) the electronic system for registration of devices referred to in Article 29(4); (b) the UDI-database referred to in Article 28; (c) the electronic system on registration of economic operators referred to in Article 30; (d) the electronic system on notified bodies and on certificates referred to in Article 57; (e) the electronic system on clinical investigations referred to in Article 73; (f) the electronic system on vigilance and post-market surveillance referred to in Article 92; (g) the electronic system on market surveillance referred to in Article 100. […] 4. The data shall be entered into Eudamed by the Member States, notified bodies, economic operators and sponsors as specified in the provisions on the electronic systems referred to in paragraph 2. The Commission shall provide for technical and administrative support to users of Eudamed. […] | Paragraphs 2 and 4: | Inhalt Zeile 2 / Spalte 4 |
Article 40: | […] | Paragraph 2: | Commission |
Article 42: | […] […] | Paragraphs 4, 7 and 8: Paragraph 10: | Commission |
Article 43: | […] | Paragraph 2: | Commission |
Article 44: | […] | Paragraph 12: The Commission should make available to the public the summaries of the reports. | Member States Commission |
Article 46: | […] (d) enter into the electronic system referred to in Article 57 information in relation to certificates of which it has required their suspension or withdrawal; (e) inform the competent authority for medical devices of the Member State in which the manufacturer has its registered place of business through the electronic system referred to in Article 57 of the certificates for which it has required suspension or withdrawal. That competent authority shall take the appropriate measures, where necessary to avoid a potential risk to the health or safety of patients, users or others. […] | Paragraph 7: | Member States |
Article 53: in conj. with: | […] […] | Paragraph 2: The required information may also be made available via a national system, provided that compliance with requirements on notification of all other notified bodies is ensured. | Commission Notified Bodies |
Article 54: | […] […] | Paragraph 3: Note on expert panels: | Commission Notified Bodies |
Article 55: | 1. A notified body shall notify the competent authorities of certificates it has granted to devices for which the conformity assessment has been performed pursuant to Article 54(1). Such notification shall take place through the electronic system referred to in Article 57 and shall include the summary of safety and clinical performance pursuant to Article 32, the assessment report by the notified body, the instructions for use referred to in Section 23.4 of Annex I, and, where applicable, the scientific opinion of the expert panels referred to in Section 5.1 of Annex IX or Section 6 of Annex X, as applicable. In the case of divergent views between the notified body and the expert panels, a full justification shall also be included. […] | Paragraph 1: | Commission Notified Bodies |
Article 56: | […] […] | Paragraph 5: Note: As soon as the functionality is available in Eudamed, the system may be used for the upload of the certificates even before the notice of full functionality of Eudamed has been published. | Notified Bodies |
Article 70 |
[…] | Paragraph 1: Paragraph 2: Paragraphs 2 and 3: | Sponsors Commission |
Article 73 | 1. The Commission shall, in collaboration with the Member States, set up, manage and maintain an electronic system: (a) to create the single identification numbers for clinical investigations referred to in Article 70(1); (b) to be used as an entry point for the submission of all applications or notifications for clinical investigations referred to in Articles 70, 74, 75 and 78 and for all other submission of data, or processing of data in this context; (c) for the exchange of information relating to clinical investigations in accordance with this Regulation between the Member States and between them and the Commission including the exchange of information referred to in Articles 70 and 76; (d) for information to be provided by the sponsor in accordance with Article 77, including the clinical investigation report and its summary as required in paragraph 5 of that Article; (e) for reporting on serious adverse events and device deficiencies and related updates referred to in Article 80. […] 3. The information referred to in point (c) of paragraph 1 shall only be accessible to the Member States and the Commission. The information referred to in the other points of that paragraph shall be accessible to the public, unless, for all or parts of that information, confidentiality of the information is justified on any of the following grounds: (a) protection of personal data in accordance with Regulation (EC) No 45/2001; (b) protection of commercially confidential information, especially in the investigators brochure, in particular through taking into account the status of the conformity assessment for the device, unless there is an overriding public interest in disclosure; (c) effective supervision of the conduct of the clinical investigation by the Member State(s) concerned. […] | Paragraph 1: b. This should take place via the respective national procedures applicable to clinical investigations. c. This should be facilitated by means of ad hoc exchange, e.g. through regular teleconferences, or collaborative platforms (for example CircaBC or other online platforms). d. This should take place via the respective national procedures applicable to clinical investigations. CI reports and the respective summary reports should be published via the use of a dedicated publicly accessible CircaBC directory. The publication process should be coordinated by the MDCG to avoid duplications. e. This should take place via the respective national procedures applicable to clinical investigations and as described in the MDCG Guidance on safety reporting in clinical investigations. (10) Paragraph 3: | Commission, Member States (communication) |
Article 74 | 1. Where a clinical investigation is to be conducted to further assess, within the scope of its intended purpose, a device which already bears the CE marking in accordance with Article 20(1), (‘PMCF investigation’), andwhere the investigation would involve submitting subjects to procedures additional to those performed under the normal conditions of use of the device and those additional procedures are invasive or burdensome, the sponsor shall notify the Member States concerned at least 30 days prior to its commencement by means of the electronic system referred to in Article 73. The sponsor shall include the documentation referred to in Chapter II of Annex XV as part of the notification. Points (b) to (k) and (m) of Article 62(4), Article 75, Article 76, Article 77, Article 80(5) and the relevant provisions of Annex XV shall apply to PMCF investigations. […] | Paragraph 1: | Sponsors |
Article 75 | 1. If a sponsor intends to introduce modifications to a clinical investigation that are likely to have a substantial impact on the safety, health or rights of the subjects or on the robustness or reliability of the clinical data generated by the investigation, it shall notify, within one week, by means of the electronic system referred to in Article 73 the Member State(s) in which the clinical investigation is being or is to be conducted of the reasons for and the nature of those modifications. The sponsor shall include an updated version of the relevant documentation referred to in Chapter II of Annex XV as part of the notification. Changes to the relevant documentation shall be clearly identifiable. […] | Paragraph 1: | Sponsors |
Article 76 | […]
| Paragraph 3-4: | Member States |
Article 77 | 1. If the sponsor has temporarily halted a clinical investigation or has terminated a clinical investigation early, it shall inform within 15 days the Member State in which that clinical investigation has been temporarily halted or terminated early, through the electronic system referred to in Article 73, of the temporary halt or early termination, providing a justification. In the event that the sponsor has temporarily halted or terminated early the clinical investigation on safety grounds, it shall inform all Member States in which that clinical investigation is being conducted thereof within 24 hours. […] […] | Paragraph 1: Paragraph 5: Paragraph 7: | Sponsors Commission |
Article 78 | 1. By means of the electronic system referred to in Article 73, the sponsor of a clinical investigation to be conducted in more than one Member State may submit, for the purpose of Article 70, a single application that, upon receipt, is transmitted electronically to all Member States in which the clinical investigation is to be conducted. 8. Where the conclusion of the coordinating Member State concerning the area of coordinated assessment is that the conduct of the clinical investigation is acceptable or acceptable subject to compliance with specific conditions, that conclusion shall be deemed to be the conclusion of all Member States concerned. Notwithstanding the first subparagraph, a Member State concerned may only disagree with the conclusion of the coordinating Member State concerning the area of coordinated assessment on the following grounds: (a) when it considers that participation in the clinical investigation would lead to a subject receiving treatment inferior to that received in normal clinical practice in that Member State concerned; (b) infringement of national law; or (c) considerations as regards subject safety and data reliability and robustness submitted under point (b) of paragraph 4. Where one of the Member States concerned disagrees with the conclusion on the basis of the second subparagraph of this paragraph, it shall communicate its disagreement, together with a detailed justification, through the electronic system referred to in Article 73, to the Commission, to all other Member States concerned and to the sponsor. […] 12. Any substantial modifications as referred to in Article 75 shall be notified to the Member States concerned by means of the electronic system referred to in Article 73. Any assessment as to whether there are grounds for disagreement as referred to in the second subparagraph of paragraph 8 of this Article shall be carried out under the direction of the coordinating Member State, except for substantial modifications concerning Sections 1.13, 3.1.3, 4.2, 4.3 and 4.4 of Chapter II of Annex XV, which shall be assessed separately by each Member State concerned. | Paragraphs 1, 8, 11 and 12: The MDCG may provide further guidance on the voluntary procedure, if deemed necessary. | Sponsors |
Article 80 | […] 2. The sponsor shall report, without delay to all Member States in which the clinical investigation is being conducted, all of the following by means of the electronic system referred to in Article 73: (a)any serious adverse event that has a causal relationship with the investigational device, the comparator or the investigation procedure or where such causal relationship is reasonably possible; (b) any device deficiency that might have led to a serious adverse event if appropriate action had not been taken, intervention had not occurred, or circumstances had been less fortunate; (c) any new findings in relation to any event referred to in points (a) and (b). The period for reporting shall take account of the severity of the event. Where necessary to ensure timely reporting, the sponsor may submit an initial report that is incomplete followed up by a complete report. Upon request by any Member State in which the clinical investigation is being conducted, the sponsor shall provide all information referred to in paragraph 1. 3. The sponsor shall also report to the Member States in which the clinical investigation is being conducted any event referred to in paragraph 2 of this Article that occurred in third countries in which a clinical investigation is performed under the same clinical investigation plan as the one applying to a clinical investigation covered by this Regulation by means of the electronic system referred to in Article 73. 4. In the case of a clinical investigation for which the sponsor has used the single application referred to in Article 78, the sponsor shall report any event as referred to in paragraph 2 of this Article by means of the electronic system referred to in Article 73. Upon receipt, this report shall be transmitted electronically to all Member States in which the clinical investigation is being conducted. Under the direction of the coordinating Member State referred to in Article 78(2), the Member States shall coordinate their assessment of serious adverse events and device deficiencies to determine whether to modify, suspend or terminate the clinical investigation or whether to revoke the authorisation for that clinical investigation. This paragraph shall not affect the rights of the other Member States to perform their own evaluation and to adopt measures in accordance with this Regulation in order to ensure the protection of public health and patient safety. The coordinating Member State and the Commission shall be kept informed of the outcome of any such evaluation and the adoption of any such measures. […] | Paragraphs 2, 3 and 4: | Sponsors (notification) |
Article 86: | […]
| Paragraph 2: Notified bodies should provide the PSURs and corresponding evaluations to the manufacturers and make them available upon request to the competent authority. | Manufacturers Notified Bodies |
Article 87: | 1. Manufacturers of devices made available on the Union market, other than investigational devices, shall report, to the relevant competent authorities, in accordance with Articles 92(5) and (7), the following: (a) any serious incident involving devices made available on the Union market, except expected side-effects which are clearly documented in the product information and quantified in the technical documentation and are subject to trend reporting pursuant to Article 88; (b) any field safety corrective action in respect of devices made available on the Union market, including any field safety corrective action undertaken in a third country in relation to a device which is also legally made available on the Union market, if the reason for the field safety corrective action is not limited to the device made available in the third country. The reports referred to in the first subparagraph shall be submitted through the electronic system referred to in Article 92. […] The reports referred to in the ninth subparagraph shall be submitted through the electronic system referred to in Article 92. […] | Paragraph 1: The current FSCA form should be used (the additional information required under the MDR may be added to the general comments section of the form).
Paragraph 9: | Member States Manufacturers |
Article 88: | 1. Manufacturers shall report, by means of the electronic system referred to in Article 92, any statistically significant increase in the frequency or severity of incidents that are not serious incidents or that are expected undesirable side- effects that could have a significant impact on the benefit-risk analysis referred to in Sections 1 and 5 of Annex I and which have led or may lead to risks to the health or safety of patients, users or other persons that are unacceptable when weighed against the intended benefits. […] | Paragraph 1: | Member States Manufacturers |
Article 89: | […] 7. After carrying out the evaluation in accordance with paragraph 3 of this Article, the evaluating competent authority shall, through the electronic system referred to in Article 92, inform, without delay, the other competent authorities of the corrective action taken or envisaged by the manufacturer or required of it to minimise the risk of recurrence of the serious incident, including information on the underlying events and the outcome of its assessment. 8. […] The manufacturer shall enter the field safety notice in the electronic system referred to in Article 92 through which that notice shall be accessible to the public. […] 9. […] The coordinating competent authority shall, through the electronic system referred to in Article 92, inform the manufacturer, the other competent authorities and the Commission that it has assumed the role of coordinating competent authority. | Paragraph 5: Paragraph 7: Paragraph 8 (third sub-paragraph): Paragraph 9: | Member States Commission, Member States Member States Manufacturers |
Article 90: | The Commission shall, in collaboration with the Member States, put in place systems and processes to actively monitor the data available in the electronic system referred to in Article 92, in order to identify trends, patterns or signals in the data that may reveal new risks or safety concerns. […] | The monitoring of data will become possible from the date when EUDAMED Vigilance module is made available. | |
Article 93: | […] 7. The competent authority which carried out the inspection shall communicate the content of the report referred to in paragraph 6 of this Article to the economic operator that has been the subject of the inspection. Before adopting the final report, the competent authority shall give that economic operator the opportunity to submit comments. That final inspection report shall be entered in the electronic system provided for in Article 100. 8. The Member States shall review and assess the functioning of their market surveillance activities. Such reviews and assessments shall be carried out at least every four years and the results thereof shall be communicated to the other Member States and the Commission. Each Member State shall make a summary of the results accessible to the public by means of the electronic system referred to in Article 100. […] | Paragraph 4: Paragraph 7: Paragraph 8: | Commission Member States |
Article 95: | […] 4. Where the economic operator as referred to in paragraph 1 does not take adequate corrective action within the period referred to in paragraph 1, the competent authorities shall take all appropriate measures to prohibit or restrict the making available of the device on their national market, to withdraw the device from that market or to recall it. The competent authorities shall notify the Commission, the other Member States and the notified body referred to in paragraph 2 of this Article, without delay, of those measures, by means of the electronic system referred to in Article 100. 6. Member States other than the Member State initiating the procedure shall, without delay, inform the Commission and the other Member States, by means of the electronic system referred to in Article 100, of any additional relevant information at their disposal relating to the non-compliance of the device concerned and of any measures adopted by them in relation to the device concerned. In the event of disagreement with the notified national measure, they shall, without delay, inform the Commission and the other Member States of their objections, by means of the electronic system referred to in Article 100. […] | Paragraphs 2, 4 and 6: | Member States |
Article 97: | […] | Paragraph 2: | Member States |
Article 98: | […] | Paragraph 2: | Member States |
Article 99: | […] | Paragraph 4: | Member States |
Article 120: | […] | Paragraph 3: | Member States |
Footnotes
(1): Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC, OJ L 117, 5.5.2017, p. 1–175.
(2): Available at: https://ec.europa.eu/growth/sectors/medical-devices/new-regulations/eudamed_en.
(3): Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU, OJ L 117, 5.5.2017, p. 176–332.
(4): Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices, OJ L 189, 20.7.1990, p. 17–36.
(5): Council Directive 93/42/EEC of 14 June 1993 concerning medical devices, OJ L 169, 12.7.1993, p. 1–43.
(7): MDCG 2020-15
(8): Eudamed2 will not be subject to any modifications.
(9): “MDR” should be added at the beginning of the CI title, indicating the sponsor and investigational devices, without risk class. Operational guidance will be issued by MDCG CIE.
(10): MDCG 2020-10/1 Safety reporting in clinical investigations of medical devices under the Regulation (EU) 2017/745.
(11): MDCG 2020-10/1 and MDCG 2020-10/2.